In this section: Latest

Documentary record · October 19, 2026 edition

Edwards SAPIEN M3 Commander M delivery systems: FDA Early Alert

EARLY ALERT

SOURCE-VERIFIED · HUMAN REVIEW PENDING

Event date

October 15, 2026

Lifecycle state

EARLY ALERT (device issue communication)

Source published

October 15, 2026

Source checked

October 20, 2026

Published on KPSGILL

Edwards reported three balloon-burst events on specific lots (67718561, 67804014, 67887043, 67887044). These carry a risk of embolization or a non-retrievable device that may need surgery. As of October 5, Edwards had reported one death and one serious injury associated with the issue. Edwards told customers on October 5 to remove the affected lots.

Does not establish

a recall classification. The alert is addressed to facilities that stock the delivery system; it gives no instructions to patients who already have a valve.

Sources

Register: the record store · record id fda-early-alert-edwards-sapien-m3-commander-2026-10-15 · SOURCE-VERIFIED · HUMAN REVIEW PENDING.

Documentary record · event date October 15, 2026 · sources checked October 20, 2026 · published October 19, 2026.

Search KPSGill.com

Search the full publication — patient guides, policy, investigations, enforcement records and research.